The Dash family of monitors provides flexibility in a single platform. With a comprehensive set of clinical parameters, Dash monitors capture vital patient measurements while sophisticated algorithms help prevent false alarms. Dash monitors are highly configurable, enabling you to add features and parameters accommodating varying acuity levels for patients across the care continuum. From presentation in the Emergency Department to surgery in the OR to recovery in the PACU, treatment in the ICU and transfer to the Stepdown unit, the Dash monitor is an excellent choice for every point of care.
Features & Specifications
The GE DASH 4000 is a portable patient monitor that has capabilities to measure ECG, NIBP, SpO2 along with many other perimeters. The Dash 4000 can connect to the GE Carescape network either wired or wireless. The GE Dash 4000 patient monitor comes with a large 10.4-inch color screen. The GE Dash Monitor also offers 12-lead 12sl, EK-Pro arrhythmia, ST-Segment analysis, and A-fib Detection. With these advanced parameters, the GE Dash 4000 can deliver personalized care for each patient’s situation. The Dash’s lightweight and durable design makes it ideal for any situation including transport, ER, Surgical Centers, and the ICU.
Unmatched adaptability at the bedside.
Modular flexibility
Revolutionary chest pain care
Early intervention in the NICU
Gold standard NIBP accuracy
Options: Agent Monitoring Dimensions
Height: 10.7 in (27.4 cm.)
Depth: 9.5 in (24.3 cm.)
Width: 11.5 in (29.3 cm.)
Weight: 5.7 kg
Display
Size: 26 cm (diagonal).
Type: active-matrix color TFT.
Resolution: 640 by 480 pixels.
Number of traces: 6 maximum.
Number of seconds/trace: 4.9 at 25 mm/sec.
Sweep speed: 6.25, 12.5, 25 mm/sec (with erase bar).
Information window: displays non-real-time information without obstructing the display of real-time information.
Display organization: prioritized by parameter.
Controls
Trim Knob® control.
Five hard keys: silence alarm, graph go/stop, function (zero all) and power on/off.
Alarms
Categories: patient status and system status.
Priority: 4 levels – crisis, warning, advisory, and message.
Notification: audible and visual.
Setting: default and individual.
Silencing: 1 minute, current alarm only.
Pause: 5 minutes in adult ICU mode, 3 minutes in neonatal ICU mode, and 5,15, or permanent pause in OR mode.
Volume: default 70%, 70 dB measured at 1 meter.
ECG
Standard leads available: I, II, III, V, aVR, aVL, and aVF. 10
lead wire cable: I, II, III, V, aVR, aVF, V2, V3, V4, V5, V6.
Leads analyzed simultaneously: I, II, III, and V (multi-lead mode).
Lead fail: identifies failed lead.
Alarms: user-selectable upper and lower heart rate limits.
Input Specifications
Voltage range: ±0.5 mV to ±5 mV.
Signal width: 40 ms to 120 ms (Q to S).
Heart rate range: 30 to 300 bpm.
Input impedance: common mode > 10 Mat 50/60 Hz, differential > 2.5 Mfrom dc to 60 Hz.
Common mode rejection: 90 dB minimum at 50 or 60 Hz
Respiration
Measurement technique: impedance variation detection.
Range respiration rate: 0 – 200 breaths per minute.
Base impedance: 100 – 1000 Ωat 52.6 kHz.
Detection sensitivity: 0.4 to 10 Ωvariation.
Waveform display bandwidth: 0.1 to 1.8 Hz (-3 dB).
Alarms: user-selectable upper and lower respiration limits, and user-selectable apnea limit.
Temperature
Number of channels: 2.
Input specifications – probe type: YSI series 400 or 700,
temperature range: 0°C to 45°C (32°F to 113°F),
resolution: ±0.1°C.
Output specifications – parameters displayed: T1, T2,
accuracy: ±0.1°C for YSI series 4000 probes, ±0.3°C for YSI series 7000 probes, independent of source.
Alarms: user-selectable upper and lower limits for T1 and T2.
Invasive Blood Pressure
Number of channels: 2 (optional).
Transducer sites: arterial, femoral artery, pulmonary arterial, central venous, right atrial, left atrial, intracranial, and special.
Transducer requirements: 5 Vdc±0.1% excitation voltage.
Transducer output: 5 μV/V/mmHg. Input specifications – range: -25 mmHg to 300 mmHg, offset: ±150 mmHg.
Output specifications – frequency response: dc to 50 Hz (-0/+2 Hz).
Zero balance range: ±150mmHg.
Zero balance accuracy: ±1mmHg.
Zero balance drift: ±1mmHg over 24 hours.
Accuracy: ±2% or ±1mmHg, whichever is greater.
Alarms: user- selectable upper and lower limits for systolic, diastolic, and mean pressures.
NIBP
Measurement technique: oscillometric.
Displayed parameters; systolic, diastolic, and mean pressures.
Systolic: 30-275 mmHg adult, 30-235 mmHg pediatric, and 30-135 mmHg neonate.
Diastolic: 10-220 mmHg adult, 10-220 mmHg pediatric, and 10-110 mmHg neonate.
MAP: 20-260 mmHg adult, 20-260 mmHg pediatric, and 20-125 mmHg neonate.
Pulse rate, as displayed in tabular trends: 30-200 bpm adult, 30-200 bpm pediatric, and 30-200 bpm neonate
Pulse Oximetry
Parameters monitored: arterial oxygen saturation (SpO2) and peripheral pulse rate (PPR).
SpO2range: 50 –100%. PPR range: 30-300 beats per minute.
SpO2 accuracy: ±2% (70-100%) and ±3% (50- 69%).
PPR accuracy: ±3 beats per minute.
Alarms: user-selectable upper and lower limits for SpO2 and PPR.
CO2
Information displayed: inspired and expired CO2 concentrations in %, mmHg, or kPa, respiratory rate, and continuous CO2 waveform.
Measurement ranges: PiCO2/FiCO2: 0 to 100 mmHg/0to13%. Pe CO2/ Fe CO2: 0 to 100 mmHg/0 to 13%;
RR: 0 to 120 breaths/min.
Accuracy: CO2: 5% of reading or ±2 mmHg, whichever is greater.
Display resolution: 1 mmHg. Rise time: less than 60 msec.
Respiration rate accuracy: ± breath/min Warranty:
For new products, please refer to the manufacturer’s website for details.
This product has been refurbished and tested to the highest quality standards by The PCB Medical. This Limited Warranty offered by PCB Medical covers defects in material or workmanship. This warranty extends to the original purchaser only and is nontransferable. The warranty begins the day the equipment is received. Please inspect your item(s) upon delivery. Any unauthorized service providers working on our equipment within the warranty period will void the warranty!
Our warranty periods are very based on product, a standard 30 day parts warranty is included with all refurbished parts unless stated otherwise on the quote/Invoice. Warranties begin from the documented date of receipt.
Our warranties do not cover any problem that is caused by: A. Conditions, malfunctions or damage not resulting from defects in material or workmanship. B. Conditions, malfunctions or damage resulting from (1) normal wear and tear, improper installation, improper maintenance, misuse, abuse, negligence, accident or alteration.
Our limited warranties are void if a product is returned with removed, damaged or tampered labels or any alterations (including removal of any component or external cover).
PCB Medical will not provide any warranty coverage unless claims are made in compliance with all terms of the controlling warranty statement included with your product and you follow proper return procedure. To request warranty service, you will need to provide: 1. The invoice or other evidence of the date and place of purchase. 2. A description of the problem. 3. Contact our Customer Service Department to receive a Return Authorization Number (RMA#)
Shipping and Returns:
Damaged freight needs to be reported within 24 hours to file a claim. No merchandise may be returned without prior written authorization by PCB Medical.
No NEW merchandise may be returned unless authorization from manufacturer which could result in a restocking fee or no return at all, and customer is responsible for shipping expenses.
Return period for products is thirty (30) days from delivery or otherwise agreed upon.
The PCB Medical will test returned products returned and can impose a customer fee if the customer misrepresents the condition of the product.
CREDIT AND RETURN SHIPPING GUIDELINES
Order cancellations (after 24 hours) and returns are charged a 45% restocking fee. Damaged goods will receive credit dependent on the condition of the product, and customer is responsible for shipping expenses.
No credit will be issued if the item is non-returnable, such as a new product, custom or special order, or in the case that the item is damaged due to improper handling/packaging by the customer.
Shipping Charges: Customers are responsible for making sure items can be delivered into or outside of their facility without issue. For example, can a 53’ foot truck access your building? Failure to accommodate delivery can result in items not being delivered, additional fees, or return of equipment.
Basic Shipping Charge is for dock to dock shipping and/or liftgate service shipping. The items shipping dock to dock and/or liftgate are skidded/palletized items. It doesn’t not include inside delivery.
Residential shipping charge is for customer’s needing delivery in a residential area.
Inside delivery is for customer’s needing their purchase delivered inside the building. The item(s) will be delivered into the lobby only. They are not delivered into rooms/suites, elevators, or stairs.
Customers are responsible for disposing of packing materials.
If you are needing white glove delivery, please contact us directly at INFO@pcbhme.com